Glass Pipette
Configuration shown · Pipette alone; pipette filler, liquid and printed class marking not shown

Laboratory & Chemical

Glass Pipette

A one-mark transfer pipette in borosilicate 3.3, calibrated to deliver at 20 °C within ISO 648 Class A limits of ±0.020 mL at 10 mL and ±0.030 mL at 25 mL.

Class A ±0.03 mL at 25 mLCalibrated to deliverBulb form for one volume
Stock status
In stock
Lead time
2–3 weeks

Final availability, dimensions, materials, assembled components and commercial terms are confirmed against the selected drawing and approved sample.

Fixed-volume delivery

Choose the accuracy class required by the method

Never pipette by mouth

A bulb or filler is part of the specification

Tip residue stays

Drain-out types are not blown out

Glass Pipette profile

A one-mark pipette for a specified transfer volume

One-mark and graduated types, delivery mode & class

Pipette type, calibration and delivery technique

A one-mark pipette delivers a fixed volume, while a graduated pipette supports variable transfers within its range. Select the form and accuracy class required by the method. Confirm the delivery mode, calibration and instructions on the actual instrument.

01

Choose one-mark or graduated delivery

A one-mark transfer pipette under ISO 648 delivers a fixed volume, with Class A limits of ±0.020 mL at 10 mL and ±0.030 mL at 25 mL. A graduated pipette under ISO 835 supports variable delivery within its range. Select the instrument and accuracy class specified by your analytical method.

02

Follow the marked delivery mode

Drain-out and blow-out pipettes use different delivery conventions. Residual liquid in a drain-out tip is accounted for during calibration and should not be forced out. Identify the delivery mode from the actual instrument markings and instructions before use rather than relying on appearance alone.

03

Keep delivery technique consistent

The delivery tip is designed for a defined drainage period. Tilting the pipette or forcing delivery changes the drainage conditions. Hold it vertically, allow free drainage, touch the tip to the receiving vessel wall and observe the specified waiting time before withdrawing it.

04

Use a bulb or pipette filler

Use a bulb or pipette filler; never pipette by mouth. Check that the stem fits the filler cone and that the pipette reaches the working depth safely without an unsuitable adapter.

05

Maintain a clean, wettable surface

Accurate delivery depends on an evenly draining liquid film. Grease and dried residue can disrupt wetting and compromise a ±0.03 mL tolerance. Specify cleaning and conditioning requirements with the analytical method.

06

Account for calibration temperature

Volumetric glassware is calibrated at a stated reference temperature. Both the liquid and glass change volume as temperature changes, so account for the working temperature when using a pipette calibrated at 20 °C. Record it where required by the method.

Application guide

Fixed-volume transfers with defined delivery conditions

A one-mark pipette has a single calibrated volume. Confirm the required capacity, accuracy class and delivery convention with the analytical method and the supplied instrument.

Gloved laboratory worker using a glass pipette above a flask

Volumetric transfer

Aliquots taken for analysis

For transferring a specified aliquot before analysis or dilution. Select the calibrated capacity and accuracy class required by the method, and control reading and delivery technique. A single graduation still requires careful meniscus alignment.

  • Confirm the calibrated volume and class
  • Fill to the mark at eye level
  • Follow the stated delivery time
Glass pipette tip above the mouth of a laboratory bottle

Delivery technique

How the last drop is treated

The treatment of residual liquid depends on the pipette’s calibration and delivery mode. Identify drain-out or blow-out instructions from the actual instrument markings and manufacturer’s guidance, then follow the stated procedure consistently.

  • Confirm drain-out or blow-out
  • Never blow out a drain-out pipette
  • Train operators on the technique
Laboratory worker using a bulb filler with a glass pipette

Training new staff

A bulb or pump, never the mouth

Use a bulb or pipette filler; never pipette by mouth. Check the stem’s fit and seal with the laboratory’s chosen filler, and train users in the instrument’s specified delivery technique.

  • Confirm the stem fits the fillers used
  • Check the seal against the bulb
  • Never permit mouth pipetting

Confirm pipette type, volume, accuracy class, filler compatibility and delivery procedure against the analytical method.

FAQ

Glass Pipette questions

Questions about pipette selection, delivery mode and technique

Should I blow out the last drop?

Only when the instrument’s delivery instructions require it. Residue in a drain-out pipette is accounted for in calibration. Confirm the marked delivery mode and manufacturer’s instructions before using a blow-out action.

Is Class A worth the money for routine work?

Choose the accuracy class required by your analytical method. Cleanliness and technique remain essential with either class: residue can disrupt drainage, and changing to Class A will not correct a poorly wetting surface. Clean and pre-rinse the pipette as the method requires.

Why does our delivered volume drift over a session?

Check for residue affecting surface wetting, changes in liquid temperature and inconsistent drainage or tip withdrawal. Standardize the rinsing, drainage and touch-off procedure before attributing the drift to the glassware.

Does the calibration hold at any temperature?

Calibration applies at the stated reference temperature, normally 20 °C. Differences in glass and liquid temperature can affect delivery, so record the working temperature and follow the method’s correction requirements.

Next step

Send us the Glass Pipette brief.

Share your product, fill volume, preferred closure and destination market. We will review the matching drawings, available configurations and sample options before preparing a quote.

  • 01Send the briefProduct, volume, closure and market
  • 02Get the optionsMatching drawings and configurations
  • 03Approve the sampleConfirm the pack before production